QCI-005 Operation of Bond Strength Tester

  • PURPOSE

  • To define a procedure for operation of the Bond/Seal Strength Tester.

  • SCOPE

  • This SOP applies to the Quality Control Department.

  • REFERENCE(S) & ATTACHMENTS

  • References:

    • In-house.
  • Attachments:

    • Nil.
  • RESPONSIBILITY

  • QC Department Head:

    • To prepare the SOP.

    • To confirm the content, correctness, and detailing of the SOP.

  • QA Head:

    • To approve the new or revised SOP.

    • To ensure implementation of the defined system.

  • DISTRIBUTION

  • Quality Assurance.

  • Quality Control.

  • DEFINITION(S) & ABBREVIATION(S)

  • Definitions:

    • Nil.
  • Abbreviations:

    • Nil.
  • PROCEDURE

  • For Bond Strength Testing:

    • Cut the specimen to 15 mm of the sample to be tested. De-laminate the layers between which the bond is to be checked.

    • Power ON the system and attach one end of the de-laminated specimen to the upper jaw (movable jaw) with the help of self-adhesive tape.

    • Now attach the other end of the de-laminated specimen to the lower (fixed jaw) roller of the equipment.

    • Ensure that the sample is not stretched between the holding ends, so that the test starts from origin.

    • Press Zero, Reset, and Start, then press the load switch. NOTE: On startup, the system will always start in Peak HOLD mode.

    • The display will show the intended force (at breakage) and the same value shall be held until the Zero key is pressed again.

  • For Sealing Strength Testing:

    • Cut the specimen to 15 mm of the sample to be tested (Aluminium foil + PVC / Alu-Alu).

    • Fix both the samples in paper and seal at the Heat Seal Tester machine at the desired temperature.

    • The same procedure applies from step 7.1.2 onwards.

  • REVISION HISTORY

+----------------+----------------+----------------+----------------+----------------+----------------+ | Revision No. | 00 | Effective | | CC No. | | | | | Date | | | | +----------------+----------------+----------------+----------------+----------------+----------------+ | Details of revision: New SOP | +-----------------------------------------------------------------------------------------------------+